Following a 2-hour incubation at 35C, the operator conducts “Read 2” in the same manner as Read 1, except the “Read 2” icon is clicked. The CultureStat software records the new readings at the same two wavelengths and compares the Read 2 values to the Read 1 values. The system determines whether the sample contains log phase bacteria based on the value differences for the two wavelengths. If both wavelengths decrease significantly, log phase bacteria are present and the sample is declared positive. A decrease in only one wavelength is usually an indication of lag phase contaminants or below threshold concentrations, and the sample is declared negative. No movement means no growth is occurring, and the sample is declared negative.